Changzhou Hengheng Electronic Instrument Co., Ltd. welcome you!
QR code
QR code
0519-83263587

HENGZHENG

INTO HENGZHENG ELECTRONICS

YOUR CURRENT LOCATION:首页 > Technical applicationWEBSITE ANNOUNCEMENT:Changzhou Hengzheng Electronic Instrument Co., Ltd. welcome you!

NEWS RECOMMENDATION

What is weight loss?

What is weight loss? There are many customers in the consul...

According to its display and ...

Moisture detector based on the principle of drying directly to ...

"One way along the way" pla...

From October 26 to 29, the Fifth Plenary Session of the 18...

An Overview of the Instrum...

At present, China's instrumentation industry is in high-speed...

FA / JA series of electron...

Place the balance on a stable, vibration-free workbench, as fa...

Electronic balance and classi...

Ultramicro electronic balance: the maximum weighing of the...

Viscometer in the use of the...

Rotational viscometer is widely used in the determination of...

CONTACT US

Address: Changzhou City, Jiangsu Province, Zhongshan City, Long Road 2188, the new gate of Science and Technology ParkContact: Miss Miss
Post Code: 213000
Tel: 0519-83263587
Fax: 0519-83268313
Website: http://www.czhz17.com

1 open question _ no display after opening
(1) There is no 220V current on the power outlet to turn on the AC outlet
(2) AC adapter error
Choose the 220V ~ AC adapter (external transformer) suitable for our work
AB-S / A8-1.4v ~ 50 / 60HZ6VA
PB-S / A9.5-20VDC6W
AB / PB / GB9.5-14.5V50 / 60HZ1.5VA
(3) AC adapter burned
Replace the AC adapter
2 open the problem _ completely no display after opening
(1) Calibration data is missing
Recalculate the electronic balance
(2) instantaneous interference effects
3 open the problem _ open completely no display
Using the balance of the school, AB-S / A, PB-S / A
Using the balance of the outside school, AB / PB / GB / SB
(1) Electronic balance placed environment bad Windproof window is not closed
Improve the electronic balance of the placement environment, close all windproof windows
(2) calibration characteristics
AB-S / A balance 1 hour after the first warm-up, do the first automatic calibration; the second environment is open after the balance 2 hours; and then maintain the power balance after 150 hours, on-time automatic calibration if necessary, you can manually trigger Calibration (internal school).
The PB-S / A balance is automatically calibrated when it is turned on and, if necessary, the calibration can be triggered manually at any time. AB / PB / GB / SB can only be outside the school.
(3) calibration error
AB / PB / GB / SB with the wrong, the external weight of the school. When entering the outside school program, the balance will appear a flashing weight value, the use of this value of the weight of the outside school.
(4) AB-S / APB-S / A balance display at the top right of the "CALL"
When the display appears "CALL", the monitor makes the balance is now working, and need to do the school. After completing the school "CALL" will automatically disappear.
4 open the problem _ completely no display after opening
(1) The electronic balance placed in the environment is too bad
Improve the environment
(2) The parameters in the electronic balance menu are not set well
After recalling the menu, use the "RESET" function to exit the menu correctly and return to the factory settings
(3) weighing the room to keep the body temperature
Minimize this human factor
(4) the temperature of the object to be weighed is not equal to the balance
Place the sample beside the balance.
(5) samples exist water absorption, water discharge, static electricity, magnetic ... ... and other characteristics.
Refer to the METTLER TOLEDO balance for proper weighing / electronic analytical balance to overcome these effects in order to properly weigh and properly use the microbalance.
5 open the problem _ open completely no display
(1) the balance placed in the environment is too bad
Environmental factors include: vibration, airflow, temperature, external magnetic field, must improve the environment, close the weighing room of the windshield, balance to work properly.
(2) The parameters in the balance menu are not set well
After the menu is transferred, use the "RESET" function to exit the menu correctly and return to the factory settings
6 open the problem _ completely no display after opening
1, the desiccant of the water and water to form a different direction of the air flow, causing the air buoyancy changes, resulting in weighing instability
The desiccant in the weighing chamber should be removed to maintain a stable weighing environment
7 open the problem _ completely no display after opening
(1) not loaded plate
After the power, first installed the correct weighing plate, and then open the balance
(2) said the disk wrong
Use the correct title plate to match the balance
(3) said the disk and the wind collision
Due to improper installation of the reasons, please find out the reasons for re-correct installation.